To enable further study and assessment of indoor inhalation exposure risk, an online apparatus enabling measurement of semi-volatile compound partitioning on household particulates was developed. An example for use of the apparatus is described using dimethyl phthalate (DMP). The system employs direct measurement by membrane introduction mass spectrometry (MIMS). The MIMS system was calibrated using known gas phase DMP concentrations produced by gravimetrically calibrated permeation devices. The quantity of DMP sorbed by particles is described first using a model particle type, a reverse-phase liquid chromatography packing material, and then with a household dust sample. In addition, the desorption of semi-volatile compounds from a household dust sample was monitored using the apparatus, and characteristic fragment ion signals for phthalate compounds were observed.


Assuntos
Poluição do Ar em Ambientes Fechados/análise , Poeira/análise , Espectrometria de Massas/métodos , Sistemas On-Line , Material Particulado/análise , Ácidos Ftálicos/análise , Adsorção , Calibragem , Cromatografia Gasosa-Espectrometria de Massas/métodos , Cromatografia Gasosa-Espectrometria de Massas/normas , Humanos , Espectrometria de Massas/normas , Sistemas On-Line/instrumentação , Sistemas On-Line/normas , Material Particulado/farmacocinética , Ácidos Ftálicos/farmacocinética